ZR-1006 Mask Particulate Filtration Efficiency and Air flow Resistance Tester

Short Description:

ZR-1006 mask particulate filtration efficiency and air flow resistance tester is applicable to test the particle filtration efficiency and airflow resistance of masks and filter materials for centers of medical device inspection, security inspection center, drug inspection center, centers of diseases control and prevention, center of textile inspection, hospitals and mask R&D manufacturers.


  • Median diameter of salt aerosol: (0.075) μm
  • Median diameter of paraffin aerosol (Optional): (0.185) μm
  • Test flow: (8~100) L/min
  • Host size(L×W×H): (Length 550 × Width 550 × Height 1630) mm
  • Power consumption: <300 W
  • Weight of host: About 90 kg
  • Product Detail

    Specification

    ZR-1006 mask particulate filtration efficiency and air flow resistance tester is applicable to test the particle filtration efficiency and airflow resistance of masks and filter materials for centers of medical device inspection, security inspection center, drug inspection center, centers of diseases control and prevention, center of textile inspection, hospitals and mask R&D manufacturers.

    Standards

    GB/T 32610-2016       Technical specification of daily protective mask

    GB/2626-2019            Respiratory protection——Non-powered air-purifying particle respirator

    GB 19082-2009          Technical requirements for single-use protective clothing for medical use

    GB 19083-2010          Technical requirements for protective face mask for medical use

    TAJ 1001-2015           PM2.5 Protective mask

    T/CNTAC 55—2020   T/CNITA 09104——2020 Civil sanitary mask

    T/CTCA 7—2019       Ordinary protective mask

    YY 0469-2011            Surgical mask

    EN 149 Respiratory protective devices - Filtering half masks to protect against particles - Requirements, testing, marking

    NIOSH 42 CFR Part 84 Respiratory Protective Devices

    Features

    High definition LCD colored touch screen makes contents more intuitive and operation easier.

    The equipped dedicated salt aerosol generators can generate aerosol in specific size and concentration.

    Equipped multiple series of special fixtures are applicable to the test of various types of masks

    Embedded long life photometer module, automatically count sampling time and prompt light path cleaning

    To prevent human intervention, it can automatically calculate filtration efficiency and test airflow resistance of masks.

    The embedded high-precision electronic flowmeter and high-performance sampling pump can insure the stability of flow. 

    Embedded compressor has the function of automatic pneumatic clamping.

    Equipped with electrostatic removal device

    The automatic infrared anti-pinch protection function to protect personnel safety.

    No aerosol leakage, strong operator protection

    Data of detection can be exported by USB disk or printed by Bluetooth printer.

    Optional oil or salt aerosol generator.

    Deliver Goods

    deliver goods Italy
  • Previous:
  • Next:

  • The Main Parameter Parameter Range Resolution Maximum Permissible Error

    Median diameter of salt aerosol

    (0.075) μm

    /

    ±0.02

    Median diameter of paraffin aerosol (Optional)

    (0.185) μm

    /

    ±0.02

    Test flow

    (8~100) L/min

    0.1 L/min ±2.0%

    Range of pressure detection

    (0~2500) Pa 0.1 Pa ±2.0%

    Range of concentration detection

    (0.001-200) mg/m3 0.001 mg/m3

    Precision of test

    / / 1%FS

    Repeatability of test

    / / 2%

    Operation temperature

    0~50℃

    Instrument noise

    <60 dB(A)

    Power supply

    AC 220 V±10%,50 Hz

    Host size(L×W×H)

    (Length 550 × Width 550 × Height 1630) mm

    Power consumption

    <300 W

    Weight of host

    About 90 kg
    Write your message here and send it to us